********************************************************************************************************* 1) Current : in Scientific notation (float) 2) Mux : 0 to 7 5 : use mux_1 6,7:OPEN 3) Mux_1 : 0 to 7 0 : 100 Ohm 1: 1k 2: 10k 5,6,7:OPEN 4) Voltage CH1 : 5) Voltage CH2 : 6) Voltage CH3 : 7) Voltage CH4 : 8) Voltage CH5 : DAC code 1 correspond to 5V/4095 = 1.22mV 9) Voltage CH6 : 10) Voltage CH7 : 11) Voltage CH8 : 12) Integration time : 1: 40us 2: 80us 13) Default value : Default value of the expected sum 14) Tolerence Min : Minimal limit of the expected sum (1mA - 1nA = +25% ; 100pA - 10pA = +100% ; 1pA = +300%) 15) Tolerence Max : Maximal limit of the expected sum (1mA - 1nA = -25% ; 100pA - 10pA = -50% ; 1pA = -200%) 16) Waiting time : Time between the setup of the currents and the measurements 17) Channel to test : 1 to 8 for one channel; 0 for all ********************************************************************************************************* 1e-12 0 7 26 21 12 14 30 16 23 15 11 0.0002048 -0.0002048 0.0008192 40 1e-11 0 7 118 117 103 111 124 102 110 102 11 0.002048 0.001024 0.004096 40 1e-10 0 7 1056 1072 1016 1073 1065 979 990 976 11 0.02048 0.01024 0.04096 40 1e-9 1 7 646 996 654 988 1030 636 1039 1018 11 0.2048 0.153600 0.256000 40 1e-8 2 7 861 832 826 827 858 841 852 833 11 2.048 1.53600 2.56000 40 1e-7 3 7 831 828 813 813 841 825 831 826 10 20.48 15.3600 25.6000 10 1e-6 4 7 830 827 813 818 834 826 835 826 10 204.8 153.6 256 10 1e-5 5 2 2645 2668 2521 2602 2676 2593 2369 2597 3 2048 1536 2560 10 1e-4 5 1 2300 2335 2294 2315 2295 2330 2362 2363 2 20480 15360 25600 10 1e-3 5 0 2120 2445 2291 2203 2606 2660 2560 2355 1 204800 153600 256000 10