********************************************************************************************************* 1) Current : in Scientific notation (float) 2) Mux : 0 to 7 5 : use mux_1 6,7:OPEN 3) Mux_1 : 0 to 7 0 : 100 Ohm 1: 1k 2: 10k 5,6,7:OPEN 4) Voltage CH1 : 5) Voltage CH2 : 6) Voltage CH3 : 7) Voltage CH4 : 8) Voltage CH5 : DAC code 1 correspond to 5V/4095 = 1.22mV 9) Voltage CH6 : 10) Voltage CH7 : 11) Voltage CH8 : 12) Integration time : 1: 40us 2: 80us 13) Default value : Default value of the expected sum 14) Tolerence Min : Minimal limit of the expected sum (1mA - 1nA = +25% ; 100pA - 10pA = +100% ; 1pA = +300%) 15) Tolerence Max : Maximal limit of the expected sum (1mA - 1nA = -25% ; 100pA - 10pA = -50% ; 1pA = -200%) 16) Waiting time : Time between the setup of the currents and the measurements 17) Channel to test : 1 to 8 for one channel; 0 for all ********************************************************************************************************* 1e-12 0 7 1 1 1 1 1 1 1 1 11 0.0002048 -0.0002048 0.0008192 40 1e-11 0 7 89 79 106 90 85 85 95 85 11 0.002048 0.001024 0.004096 40 1e-10 0 7 1013 973 1070 1005 980 985 1060 977 11 0.02048 0.01024 0.04096 40 1e-9 1 7 597 590 600 645 640 670 1051 895 11 0.2048 0.153600 0.256000 40 1e-8 2 7 817 812 810 825 815 817 812 809 11 2.048 1.53600 2.56000 40 1e-7 3 7 800 796 794 805 800 799 799 797 10 20.48 15.3600 25.6000 10 1e-6 4 7 829 797 799 810 802 807 797 798 10 204.8 153.6 256 10 1e-5 5 2 2794 2767 2400 2606 2714 2400 2700 2800 3 2048 1536 2560 10 1e-4 5 1 2400 2400 2400 2400 2400 2400 2400 2700 2 20480 15360 25600 10 1e-3 5 0 3000 3000 3000 3000 3000 3000 3000 3000 1 204800 153600 256000 10